The basics

How is laser light made, and what determines what it does in the skin?

A laser emits light of a single wavelength, and that wavelength determines which substance in the skin absorbs the light and how deeply it penetrates. How much energy the target receives, and how quickly, depends on the settings: fluence, pulse duration, spot size, pulse shape, repetition rate and cooling. This page explains what each of them does, why a laser's design determines which settings it can deliver, and how laser light differs from IPL and LED.

What light does in the skin

Reflection, scattering, absorption

A small proportion of the light is reflected at the skin surface. The rest enters the skin and is scattered there, in the dermis mainly by the collagen fibres [1], until it is absorbed somewhere or leaves the skin again. Only absorbed light has an effect: there its energy is converted, for the lasers on this page mostly into heat.

Chromophores

The molecules that absorb light are called chromophores. For the lasers used in aesthetic practice, three matter: melanin, in the epidermis, the hair and pigmented lesions; haemoglobin, in the blood; and water, which is present in all tissue. Each has its own absorption spectrum, absorbing some wavelengths strongly and others hardly at all [2]. Tattoo ink absorbs light too; which wavelength targets which ink colour is covered under picosecond and Q-switched lasers.

The chart overlays the four spectra, with the wavelengths of our lasers as vertical lines. Each line links to the page for that technology. The vertical axis is logarithmic: each step is a factor of ten.

How strongly melanin, haemoglobin and water absorb light

Line chart with logarithmic axes: the absorption coefficient of melanin, oxy- and deoxyhaemoglobin and water from 400 to 11,000 nm, with vertical lines at the wavelengths of our lasers. Melanin absorbs most at short wavelengths and less and less towards the infrared. Haemoglobin peaks in the green and yellow. Water barely absorbs visible light and absorbs strongly in the mid-infrared. The table below the chart gives the values at each wavelength.

  • Melanin (melanosome)
  • Oxyhaemoglobin (blood)
  • Deoxyhaemoglobin (blood)
  • Water

Our wavelengths

Show the values as a table

Absorption coefficient µa in cm⁻¹ at the wavelengths of our lasers. A dash means the source gives no value at that wavelength.

  • 532 nm · Vascular lasers

    Melanin (melanosome)
    555
    Oxyhaemoglobin (blood)
    225
    Deoxyhaemoglobin (blood)
    187
    Water
    0.000429
  • 577 nm · Vascular lasers

    Melanin (melanosome)
    418
    Oxyhaemoglobin (blood)
    270
    Deoxyhaemoglobin (blood)
    201
    Water
    0.000886
  • 755 nm · Alexandrite

    Melanin (melanosome)
    164
    Oxyhaemoglobin (blood)
    2.6
    Deoxyhaemoglobin (blood)
    8.5
    Water
    0.0263
  • 1064 nm · Nd:YAG

    Melanin (melanosome)
    49.7
    Oxyhaemoglobin (blood)
    4.52
    Deoxyhaemoglobin (blood)
    2.02
    Water
    0.151
  • 1340 nm · Nd:YAP

    Melanin (melanosome)
    –
    Oxyhaemoglobin (blood)
    3.5
    Deoxyhaemoglobin (blood)
    3.7
    Water
    2.7
  • 1550 nm · Non-ablative

    Melanin (melanosome)
    –
    Oxyhaemoglobin (blood)
    10.3
    Deoxyhaemoglobin (blood)
    10.5
    Water
    10.8
  • 1927 nm · Thulium

    Melanin (melanosome)
    –
    Oxyhaemoglobin (blood)
    81.5
    Deoxyhaemoglobin (blood)
    76.2
    Water
    123
  • 2940 nm · Erbium

    Melanin (melanosome)
    –
    Oxyhaemoglobin (blood)
    –
    Deoxyhaemoglobin (blood)
    –
    Water
    12,000
  • 10,600 nm · CO₂

    Melanin (melanosome)
    –
    Oxyhaemoglobin (blood)
    –
    Deoxyhaemoglobin (blood)
    –
    Water
    846
Sources: water, Segelstein 1981 via refractiveindex.info (CC0 1.0); blood at 45% haematocrit, Bosschaart et al. 2014 (CC BY 4.0), from 400 to 1995 nm, from about 1400 nm mostly the water in the blood; melanin, Jacques’ approximation for a single melanosome based on Jacques & McAuliffe 1991, from 400 to 1200 nm only. Real melanosomes vary up to tenfold.

Read each curve from left to right: it shows how strongly that single chromophore absorbs at each wavelength. Melanin, for example, absorbs about eleven times less at 1064 nm than at 532 nm. Comparing the curves with each other only goes so far, because each applies to a different amount of substance: melanin to the inside of a melanosome, haemoglobin to whole blood and water to pure water. How much light the epidermis absorbs in total depends on how many melanosomes it contains, and so on the skin type. Above about 1400 nm, absorption by whole blood is due mainly to its own water.

How deeply light penetrates

How deeply light penetrates depends on how much of it is absorbed and how much is scattered on the way. Short wavelengths are scattered more strongly than long ones [2], and melanin and blood absorb them more; green light therefore penetrates less deeply than near-infrared light. Between about 600 and 1300 nm, the chromophores in the skin absorb relatively little. This range is called the optical window, and it is where light penetrates deepest [1].

Above 1300 nm, water absorbs far more strongly, with peaks around 1450, 1940 and 2940 nm, and the penetration depth falls. In pure water, about two thirds of 2940 nm light (Er:YAG) is absorbed within less than a micrometre; for 10,600 nm (CO₂), the same fraction takes more than ten micrometres [3].

The wavelength thus sets which chromophore absorbs the light, how deeply it penetrates and how much energy each photon carries: the shorter the wavelength, the more. For heating, the energy per photon does not count, only the total energy; photon energy matters again in the section on frequency doubling below. How much energy the target receives, over what area and in how much time, is governed by the settings.

What the settings control

Fluence

Fluence is the energy of one pulse divided by the area of the spot, in joules per square centimetre (J/cm²). At the same pulse energy, a small spot therefore gives a much higher fluence than a large one. Halve the diameter and the area falls to a quarter, so the fluence quadruples. Conversely, a larger spot needs more energy per pulse for the same fluence, and that limits which combinations of spot size and fluence a device can deliver.

To heat or break up a target, the fluence has to be above a threshold. If it is too far above it, the surrounding tissue also receives too much heat. Where those limits lie depends on wavelength, pulse duration, spot size and skin, and the treatment protocols specify them.

In fractional treatment, the beam is split into a grid of microscopic zones. Within each zone the fluence is high, and the skin between the zones stays intact [4]. What that means for ablative and non-ablative lasers is explained in the technology overview.

Pulse duration and peak power

Pulse duration is how long a single pulse lasts, and it controls two things. The first is whether the heat stays in the target. Every structure that has absorbed light loses its heat to its surroundings, and a small structure does so faster than a large one. The time in which it loses about half of that heat is called the thermal relaxation time. If the pulse lasts no longer than that, heating stays largely confined to the target and the surrounding tissue is spared. That is the principle of selective photothermolysis [5]. A thick blood vessel therefore takes longer to cool than a thin one, and tattoo ink particles lose their heat within ten nanoseconds [6].

For hair removal, this rule has been extended. The pigment sits in the hair shaft, but the actual target is the stem cells in the follicle wall next to it. The heat must therefore be able to travel from the shaft to those cells, and for that the pulse may be longer than the thermal relaxation time of the shaft itself [7].

The second is peak power: the energy divided by the time over which it is released. As an illustration, not a treatment setting: a pulse of 0.5 joules lasting 10 milliseconds has a power of 50 watts. The same 0.5 joules in 5 nanoseconds is 100 megawatts, and in 500 picoseconds 1 gigawatt. The energy is the same each time, but between the longest and the shortest pulse it is released twenty million times faster.

Millisecond pulses heat a whole structure, such as a blood vessel or a hair follicle. With nanosecond pulses, a particle heats up so quickly that it breaks apart, through the heat and through the pressure wave this creates [8]. Picosecond pulses shift that balance further towards the mechanical effect, with less heat for the surrounding tissue. How that works and what it means clinically is explained under picosecond and Q-switched lasers.

Spot size

Once inside the skin, the light fans out through scattering. At the edge of the beam, some of it is lost sideways, outside the treated area. With a small spot, the edge makes up a large part of the beam, and the fluence falls quickly with depth. With a large spot, light scattered away from one point makes up the shortfall at another, and the beam keeps more energy at depth. In a computer model of port-wine stain treatment, a blood vessel in the dermis absorbed 2.5 times more energy at the same fluence when the beam diameter was increased from 0.2 to 1 mm or more, while the epidermis received the same amount [9].

Pulse shape and beam profile

Two pulses with the same energy and the same duration can still act differently, because the energy is distributed differently over time and across the spot.

Pulse shape describes how the energy is distributed over time. In some sources the power peaks right at the start and then falls off; others keep it roughly constant over the whole pulse, known as a square pulse [10]. With a peaked pulse, the tissue briefly receives a much higher power than the average suggests. A pulse can also be delivered as a train of shorter sub-pulses. Stated and delivered values can differ: in measurements on nineteen IPL devices, the effective pulse duration of several was much shorter than stated [10]. How pulse shape changes the spectrum of IPL is explained on the IPL page.

Beam profile describes how it is distributed across the spot. In a Gaussian profile, with a hot centre, the fluence is highest in the middle and falls off towards the edge. In a flat profile, also called top-hat, it is nearly the same across the whole spot. Measurements on a pulsed dye laser showed such a fall-off from centre to edge [11]. The fluence a device displays is an average over the spot; with a hot centre, the edge of each spot receives less than that value. How far successive spots overlap is set out in the treatment protocols.

Repetition rate and heat build-up

The repetition rate is the number of pulses per second, in hertz (Hz). It matters once successive pulses hit the same area. If the next pulse arrives before the tissue has cooled, it starts from a higher temperature and the heat adds up. This is called pulse stacking. When it happens unintentionally, through rapid repetition or too much overlap, the risk of pigmentary or textural changes rises, especially in darker skin types [12].

The in-motion technique for hair removal builds up heat deliberately. The handpiece is moved over an area in repeated passes, and the temperature of the follicles rises gradually.

Cooling

Light that has to reach a deeper target first passes through the epidermis, where melanin absorbs part of it on the way. Cooling the skin surface lets the epidermis tolerate more light before it is damaged [13]. There are three common methods:

  • Contact cooling. A chilled window, often made of sapphire, rests on the skin during the pulse, as in the handpiece of the IPL SQ.
  • Cryogen spray cooling. Just before the pulse, the device sprays a short burst of cryogen onto the skin. Because the burst lasts only tens of milliseconds, cooling is largely confined to the epidermis and the deeper vessels are barely cooled [14]. The DermaV has this cooling built in.
  • Cold air. A stream of cold air cools the skin before, during and after the pulse, for example with the Cryo 7.

In a computer model, spray and contact cooling gave the same cooling profile in the skin at the same surface temperature. How deep the cooling reaches depends on how long it is applied [13].

Which of these settings a device can deliver depends on how it generates its light.

How laser light is made

The resonator

Laser stands for light amplification by stimulated emission of radiation. Every laser has three parts: a medium that can amplify light, a pump source that gives the medium energy, and two mirrors, one at each end. The pump source creates a population inversion: more particles (atoms, molecules or, in a semiconductor, electrons) in the excited state than in the lower state to which they return on emission. When a photon of the right wavelength passes, such an excited particle releases a second photon identical to the first. This is called stimulated emission. The light travels back and forth between the mirrors and grows stronger with each pass. One mirror lets a small part through, and that part is the beam that leaves the device.

The result is monochromatic light, its wavelength set by the medium, that is also coherent, with all its waves in phase [15]. Because it stays in a narrow beam, it can be carried to the handpiece through a fibre or an articulated mirror arm.

Types of source

A laser is usually named after its medium. In aesthetic practice there are five kinds:

  • Crystal or glass. A dopant in a host crystal generates the light: neodymium in YAG (Nd:YAG, 1064 nm) or in YAP (Nd:YAP, 1340 nm), erbium in YAG (Er:YAG, 2940 nm), chromium in alexandrite (755 nm). Erbium can also be doped into glass, as in the 1550 nm Er:glass laser. Such a medium is pumped by a flashlamp or by diodes.
  • Gas. In a CO₂ laser, the medium is a gas mixture containing carbon dioxide, pumped by an electrical discharge.
  • Dye. The medium is an organic dye. The best known is the pulsed dye laser at around 585–595 nm, one of the most widely used vascular lasers [15], and it is pumped by a flashlamp [11]. We do not supply a conventional pulsed dye laser; the Hollywood Spectra does have dye handpieces for 585 and 650 nm.
  • Fibre. The medium is a doped glass fibre, pumped by diodes. That is how the 2910 nm erbium fibre laser works, and also the 1927 nm thulium laser.
  • Semiconductor. In a diode laser, a semiconductor chip converts electrical current directly into laser light. In an optically pumped semiconductor laser (OPSL), diodes pump a semiconductor chip that sits in an external resonator [16], and the light is frequency-doubled inside that resonator. FlexSys Laser takes any two of three modules: 532 nm OPSL, 577 nm OPSL or a 1550 nm diode laser.

Q-switching

Without Q-switching, a pulsed laser emits light for as long as the pump source supplies energy, and the pulse lasts roughly as long as the pump pulse. A Q-switch is a fast optical switch inside the resonator, such as a Pockels cell. While it is closed, light cannot travel back and forth between the mirrors. The pump meanwhile builds up an ever larger population inversion, and that stored energy stays in the medium. When the switch opens, it is released all at once, in a pulse of a few nanoseconds with a very high peak power [8].

A continuous-wave source can also be gated into short pulses, with a shutter or by switching it on and off quickly. Such a pulse, however, has hardly more power than the source delivers continuously. By way of comparison: a 10-watt source delivers 50 nanojoules in a 5-nanosecond pulse, ten million times less than the 0.5 joules in the pulse duration example. Only when the medium stores energy and releases it all at once does a nanosecond pulse hold enough energy to break up pigment and ink. Picosecond lasers deliver even shorter pulses, of less than a nanosecond; see picosecond and Q-switched lasers.

From 1064 to 532 nm with KTP

Green 532 nm laser light usually does not come from a medium of its own. An Nd:YAG crystal produces 1064 nm, and a non-linear crystal in the beam, often potassium titanyl phosphate (KTP), converts every two photons of 1064 nm into one photon with twice the energy. Twice the energy per photon means half the wavelength: 532 nm [8]. This works with the short pulses of a Q-switched laser as well as with long pulses [8].

FlexSys Laser's 532 and 577 nm output is also frequency-doubled, but from a semiconductor chip rather than an Nd:YAG crystal; the difference is explained under vascular lasers.

Same wavelength, different source

Because the source determines which pulses are possible, two devices with the same wavelength can do very different things. 532 nm is a good example.

DermaV delivers 532 nm in millisecond pulses. These match the thermal relaxation time of a blood vessel, and DermaV is first and foremost a vascular laser. Epidermal pigmented lesions such as lentigines are also treated with long-pulsed 532 nm [17]; the vascular lasers page discusses this further.

PicoPlus delivers the same wavelength in picosecond pulses, Hollywood Spectra in Q-switched nanosecond pulses. Those pulses break up particles: red, orange and yellow tattoo ink, and pigment granules. A millisecond pulse cannot, because the heat escapes from the particle before it breaks apart.

Laser, IPL and LED

IPL and LED also work with light, but that light is produced differently and behaves differently in the skin.

An IPL device has no resonator. A flashlamp emits light across a broad band, and a filter determines which part of that spectrum reaches the skin. The light is neither coherent nor collimated. The principle is the same as with a laser, though: a chromophore absorbs the light and is heated. Because the band is broad, several chromophores absorb energy at once, and the heating is less selective [18]. What that means when pigment and redness occur together is covered on the IPL page.

An LED is a semiconductor that emits light in a narrow band, without a resonator and at a much lower power per unit area than a laser or IPL. It does not noticeably heat the tissue. The light acts on different chromophores from those targeted by lasers, including cytochrome c oxidase in the mitochondria and porphyrins [19]. LED is used as an adjunct to another treatment.

Advice

What fits your practice?

A single page cannot fully cover the choice between technologies. It depends on your patients and their skin types, the indications you want to treat, the recovery time your patients will accept and the equipment you already have. Our clinical and sales specialists have years of experience with that choice and are glad to help you make it.

Already working with one of our devices? You will find the treatment protocols and guidelines on the customer portal, and we go deeper into technique in the training after delivery and at our Academy events.

This is an indicative overview to explain how technologies differ. Actual settings are governed by the treatment protocols and guidelines.

References

  1. Anderson RR, Parrish JA. The optics of human skin. J Invest Dermatol. 1981;77(1):13–19. doi:10.1111/1523-1747.ep12479191
  2. Jacques SL. Optical properties of biological tissues: a review. Phys Med Biol. 2013;58(11):R37–R61. doi:10.1088/0031-9155/58/11/R37
  3. Hale GM, Querry MR. Optical constants of water in the 200-nm to 200-µm wavelength region. Appl Opt. 1973;12(3):555–563. doi:10.1364/AO.12.000555
  4. Manstein D, Herron GS, Sink RK, Tanner H, Anderson RR. Fractional photothermolysis: a new concept for cutaneous remodeling using microscopic patterns of thermal injury. Lasers Surg Med. 2004;34(5):426–438. doi:10.1002/lsm.20048
  5. Anderson RR, Parrish JA. Selective photothermolysis: precise microsurgery by selective absorption of pulsed radiation. Science. 1983;220(4596):524–527. doi:10.1126/science.6836297
  6. Reiter O, Atzmony L, Akerman L, Levi A, Kershenovich R, Lapidoth M, Mimouni D. Picosecond lasers for tattoo removal: a systematic review. Lasers Med Sci. 2016;31(7):1397–1405. doi:10.1007/s10103-016-2001-0
  7. Altshuler GB, Anderson RR, Manstein D, Zenzie HH, Smirnov MZ. Extended theory of selective photothermolysis. Lasers Surg Med. 2001;29(5):416–432. doi:10.1002/lsm.1136
  8. Patil UA, Dhami LD. Overview of lasers. Indian J Plast Surg. 2008;41(Suppl):S101–S113. doi:10.1055/s-0039-1700481
  9. Keijzer M, Pickering JW, van Gemert MJ. Laser beam diameter for port wine stain treatment. Lasers Surg Med. 1991;11(6):601–605. doi:10.1002/lsm.1900110616
  10. Ash C, Town G, Bjerring P. Relevance of the structure of time-resolved spectral output to light-tissue interaction using intense pulsed light (IPL). Lasers Surg Med. 2008;40(2):83–92. doi:10.1002/lsm.20596
  11. Dinehart SM, Flock S, Waner M. Beam profile of the flashlamp pumped pulsed dye laser: support for overlap of exposure spots. Lasers Surg Med. 1994;15(3):277–280. doi:10.1002/lsm.1900150307
  12. Harnchoowong S, Vachiramon V, Jurairattanaporn N. Cosmetic considerations in dark-skinned patients. Clin Cosmet Investig Dermatol. 2024;17:259–277. doi:10.2147/CCID.S450081
  13. Zenzie HH, Altshuler GB, Smirnov MZ, Anderson RR. Evaluation of cooling methods for laser dermatology. Lasers Surg Med. 2000;26(2):130–144. doi:10.1002/(SICI)1096-9101(2000)26:2<130::AID-LSM4>3.0.CO;2-J
  14. Nelson JS, Milner TE, Anvari B, Tanenbaum BS, Kimel S, Svaasand LO, Jacques SL. Dynamic epidermal cooling during pulsed laser treatment of port-wine stain: a new methodology with preliminary clinical evaluation. Arch Dermatol. 1995;131(6):695–700. doi:10.1001/archderm.1995.01690180071013
  15. Husain Z, Alster TS. The role of lasers and intense pulsed light technology in dermatology. Clin Cosmet Investig Dermatol. 2016;9:29–40. doi:10.2147/CCID.S69106
  16. Kuznetsov M, Hakimi F, Sprague R, Mooradian A. High-power (>0.5-W CW) diode-pumped vertical-external-cavity surface-emitting semiconductor lasers with circular TEM₀₀ beams. IEEE Photonics Technol Lett. 1997;9(8):1063–1065. doi:10.1109/68.605500
  17. Chan HH, Fung WK, Ying SY, Kono T. An in vivo trial comparing the use of different types of 532 nm Nd:YAG lasers in the treatment of facial lentigines in Oriental patients. Dermatol Surg. 2000;26(8):743–749. doi:10.1046/j.1524-4725.2000.00039.x
  18. Babilas P, Schreml S, Szeimies RM, Landthaler M. Intense pulsed light (IPL): a review. Lasers Surg Med. 2010;42(2):93–104. doi:10.1002/lsm.20877
  19. Jagdeo J, Austin E, Mamalis A, Wong C, Ho D, Siegel DM. Light-emitting diodes in dermatology: a systematic review of randomized controlled trials. Lasers Surg Med. 2018;50(6):613–628. doi:10.1002/lsm.22791